6-21
Performance Tests
10. Input Noise
This test measures the lock-in input noise.
Setup
Connect a 50Ω termination to the A input. This grounds the input so the lock-in's own noise is measured.
Procedure
1) {PRESET} (Turn the lock-in off and on with the [Setup] key pressed)
2) Press the keys in the following sequence:
[Sensitivity Down]
Set the sensitivity to 100 nV.
[Channel 1 Display]
Set the Channel 1 display to X Noise.
3) Wait until the reading of Channel 1 stabilizes. Record the value of Channel 1.
4) This completes the noise test. Enter the results of this test in the test record at the end of this section.